A Market Standing Certificate (MSC) is issued by the Indian licensing authority to show that a licensed medical device has been on sale in India. Government hospitals and tender bodies usually ask for it, often alongside a Non-Conviction Certificate (NCC). Medfins prepares the MSC application, the sales evidence and the undertaking so the certificate is ready before your tender closes.
Who issues a Market Standing Certificate
CDSCO’s guidance document on MSC and NCC for licensed medical devices splits applications by licence:
- Class C and Class D devices made under a manufacturing licence, and devices of any class under an import licence: apply to the Central Licensing Authority.
- Class A and Class B devices made under a manufacturing licence: apply to the State Licensing Authority.
Applications are made online through the Medical Devices portal (cdscomdonline.gov.in). Since April 2025, CDSCO has also auto-generated MSC and NCC certificates for devices licensed by the Central Licensing Authority through the same online system.
Documents for an MSC application
- Covering letter
- Application for MSC
- Copy of the valid manufacturing or import licence, with the retention application file number where applicable
- Sales evidence: sales details certified by a chartered accountant (manufacturers), or Bills of Entry or CA-certified sales details (importers)
- Legal undertaking on Rs. 100 registered stamp paper, recently notarised, stating that no action or conviction has arisen from adverse events, market complaints or Not of Standard Quality (NSQ) reports for the applied products
The sales evidence must cover the last three years from the date of application, or from the date the licence was granted if that is later, for every product applied for. Every brand and model number on the application should appear in the sales details.
How long a Market Standing Certificate is valid
An MSC is valid for one year from issue, or until the licence expires if that is earlier, provided the licensing authority has not started regulatory action on the products. A Free Sale Certificate, by contrast, stays valid as long as the manufacturing licence. If you bid for tenders every year, plan a fresh MSC each year.
Where MSC applications go wrong
- Sales records that do not name the model. Invoices that list only a product family leave the authority unable to tie sales to each applied model.
- A licence too new for the market-standing period. The sales period runs from the licence grant date when that is later, so a recently licensed device has a shorter record.
- A stale undertaking. The undertaking speaks as of its date; one notarised for an earlier application invites a query.
- The wrong authority. Class A and B manufactured devices go to the State Licensing Authority; Class C and D, and all imported devices, go to the Central Licensing Authority.
Related: Free Sale Certificate for medical devices · Neutral code for export · MD-42 registration · CDSCO manufacturing licence
Need an MSC for a tender? Contact our team with your licence number, product list and tender deadline.